The MIFETM
system
|
Home UTas Biophysics Lab MIFE user group Overview MIFE theory Hardware Software Purchasing |
Methodological issues in flux measurementThe cornerstone of each electrophysiological technique is microelectrode fabrication. For MIFE measurements, liquid membrane type ion-selective microelectrodes are used. Specific details of their fabrication and calibration are given in our publications (e.g. Shabala et al. 1997; Shabala and Shabala 2002; Shabala et al. 2006). Briefly, electrodes are pulled from non-filamentous borosilicate glass capillaries to tip diameter ~1 µm. Electrode blanks are then silanized with tributylchlorosilane (90796, Fluka Chemicals) to make their surface hydrophobic. Dried and cooled electrode blanks are stored under cover and may be used over several weeks. To make the electrode, the electrode tip is first broken slightly to achieve the required diameter (typically 2-3 µm). It is then backfilled with an appropriate solution (see Shabala et al (2006) for more details) and finally front-filled with the appropriate liquid ion exchanger (LIX). The column length of the LIX in the prepared electrode is usually 50 to 200 µm. Immediately after filling, electrodes are immersed in solution and kept there until use (up to 8-10 hours). In our laboratory, we routinely use the MIFE system to measure net fluxes of H+, Ca2+, K+, Na+, Cl-, Mg2+, NH4+, NO3-, and Cd2+ from various systems: higher plant tissues and protoplasts; animal tissues (e.g. muscles); fungi; algae; protists; yeasts; bacterial monolayers and biofilms. Several more ions (e.g. Zn2+, Cu2+, Cs+, Pb3+, SO4-2) also can be measured using commercially available ionophores. We also measure O2 fluxes. Despite the same principle being used for measurements of fluxes of each of these ions, there are some “specific features”, related to fabrication, calibration, and use of ion-selective microelectrodes to measure fluxes of a specific ion. The most critical ones are mentioned here. 1.
Basic electrode characteristics. As a rule, for most physiological conditions,
electrode characteristics are expected to be linear. Accordingly, each electrode
is calibrated in a set of three known standards, covering the range of
concentrations expected to be found in the experiment. The average responses of
electrodes are about 53-54 mV/decade for monovalent ions, and 27-28 mV/decade
for divalent ions, with a correlation R >0.999. If measurements are made at
very low concentrations, non-linearity requires that more than 3 standards
should be used. 2.
Electrode “conditioning”. Most of the prepared microelectrodes can be
used immediately after preparation, while others (e.g. H+ and Cl-)
need some conditioning time (~1h) to ensure a stable response. 3.
Responsiveness. During MIFE measurements, the electrodes are moved back and forth at
5-sec intervals. For accurate flux calculations, the LIX must “settle” at each
position. From practical experience, settling is achieved quicker if the LIX
column length is relatively short. However, in this case there is a danger of a
gradual leak of LIX out of the tip and electrode loosing its selectivity. The
compromise is achieved by optimizing the amount of silane used for electrode
fabrication and the amount of LIX used for electrode filling. 4.
Effect of ionic strength. Variations in ionic strength of solutions
might significantly affect characteristics of ion selective electrodes and result
in inaccurate estimates of ionic concentrations and, ultimately, net ion
fluxes. The actual concentration (and, thus, flux) is overestimated for
solutions with ionic strength lower than that of the standard, and is
underestimated vice versa. For many ions, such a difference may be as
big as a factor of 2 in a physiologically relevant range of concentrations
(e.g. Na+ levels 200 mM). More details are available in Shabala et al. (2006). 5.
Temperature. From general knowledge, it was expected that both the slope and the
intercept of the calibration curve for ion-selective electrodes might be
slightly affected by changing temperatures. The theory shows that the Nernst
slope is proportional to the Kelvin temperature. This effect is automatically
covered if calibration and measurements are done at the same temperature. The
effects of such temperature changes during an experiment are still open to
question, it is unknown how significant these changes are and whether they
should be taken into account. Our experiments in the 4 to 40 oC
range suggested that in all cases the Nernst slope remained > 50 mV/decade,
and the maximum inaccuracy in flux calculation did not exceed 6% (Shabala et al. 2006). Thus, no practical actions
are needed. However at temperatures above 32 oC, the LIX often
became very “noisy”. Thus, it is highly recommended that the performance of
specific LIX should be tested at high temperatures if these are to be applied
in an experiment, to ensure that the signal to noise ratio is acceptable, and
that net flux responses can be distinguished from background noise. 6.
ignal to noise ratio. Due to the thermal electron noise in electrodes, there is some
theoretical “lower limit” on the magnitude of the flux that can be measured
against the background noise (Ryan et al. 1990). There are two practical
measures to overcome this problem and to improve the sensitivity of the flux
measurements. One is to increase the travel range of the electrode (making
voltage changes larger), and another one is to increase electrode tip diameter.
For more details, see Shabala et al. (2006). 7.
Confounding effect of inhibitors. Pharmacological experiments are frequent in
plant electrophysiology. However, many of the channel blockers and metabolic
inhibitors routinely used in patch-clamp experiments may significantly affect
LIX characteristics. For example, even micromolar concentrations of CCCP
completely “killed” Ca2+ LIX, reducing electrode slope from 27 to
< 3 mV/decade. Therefore, a rigorous test of LIX performance in the presence
of inhibitors should be undertaken first.
|
Maintained by Ian Newman. Date . © University of Tasmania.